Why it matters

A nebuliser turns liquid into a mist you breathe straight into your lungs. Anything in the bowl goes with it.
Cleaning is also what keeps the machine working. Most nebulisers that seem to be failing are not failing, they are clogged, and the cause is nearly always medication residue left to dry in the bowl.
After each day's use
- Take the bowl apart as far as the instructions allow.
- Wash the bowl, the mouthpiece and the mask in warm soapy water.
- Rinse thoroughly. Unlike a spacer, nebuliser parts are rinsed.
- Shake off the excess and lay the parts on a clean cloth.
- Let everything air dry completely before reassembling.
Residue dries hard, and hard residue narrows the jet the mist is produced through. That is the mechanism behind "it takes twice as long as it used to".
Never wash the tubing
This one catches people out, because washing everything feels more thorough.
Water inside the tubing cannot be dried out, and it will sit there. Damp tubing is both a hygiene problem and a performance problem.
If there is moisture inside the tube, which is normal condensation, run the unit with the tubing attached and no medication in the bowl until it clears. That is the correct method.
If the tubing is discoloured, hardened, split or kinked, replace it.
The filter
Air filters protect the pump. They are consumable.
- Replace them at the interval stated in your instructions
- Replace them sooner in a dusty house, or with pets
- Never run the unit without a filter fitted
A blocked filter makes the pump work harder, which is how a machine ends up needing a repair rather than a filter.
How often to replace each part
Follow the intervals in your own instructions. As a general guide:
| Part | When to replace |
|---|---|
| Medication bowl | When output drops or treatments take longer. It is the first part to wear, approx 3 months of use. |
| Tubing | When discoloured, hardened, split or kinked |
| Mask or mouthpiece | When it perishes, hardens or stops sealing. Children's masks wear fastest, approx 6 months of use. |
| Air filter | At the stated interval, sooner in a dusty home |
Replacing the bowl is the single most effective thing you can do for a machine that has become slow. It is also the cheapest.
If more than one person uses it
In a clinic, or where a machine is shared between patients, cleaning between patients is a different standard from household cleaning. Follow the manufacturer's instructions for disinfection, and use separate kits per patient.
A nebuliser kit supplied with a machine is for one person.
Storing it
- Reassemble only when every part is completely dry
- Keep it somewhere clean and dry, not in a bathroom
- Keep the air inlet clear and the unit off carpet when running
- If it is stored between seasons, clean and dry it fully before putting it away, and check the bowl and filter before using it again
A machine that has sat in a cupboard for a year with a used bowl in it is the one that disappoints you on the night you need it.
Frequently asked questions
Can I put nebuliser parts in the dishwasher? Only if the instructions supplied with your device say so. Many parts cannot take the heat.
Can I boil the parts to sterilise them? Check your instructions. Heat distorts some components.
Why is my nebuliser taking longer than it used to? Usually the bowl, then the filter. Replace the bowl first.
There is water in the tubing. Is that a fault? No, it is condensation. Run the unit with the tubing attached and no medication until it clears.
How long does a medication bowl last? It depends on how often it is used. Judge it by performance: when treatments take longer, replace it.
Always follow the instructions for use supplied with your device, and your doctor's or pharmacist's advice on your medication.